Measurement quality#

Every measurement is off by some amount. To trust a result, you need to know by how much.

Accuracy, precision and resolution#

These three words describe different things, and a measurement can be good at one and poor at another.

  • Accuracy is how close a reading is to the true value.

  • Precision is how close repeated readings of the same value are to each other.

  • Resolution is the smallest change a reading can show.

A pressure gauge that always reads 2 psig high is precise, but not accurate. A 16-bit module reading a pressure transducer can show steps of 0.002 psig, while the transducer itself is accurate only to ±0.5 psig. The worked example below shows where those numbers come from.

Kinds of error#

Error

What it does

How to deal with it

Offset, or zero error

Shifts every reading by the same amount

Calibrate it out, or check the reading at zero

Gain, or span error

Scales every reading by a small percentage, so it grows with the reading

Calibrate it out

Nonlinearity

Bends the response away from a straight line

Calibrate at several points, or accept it

Hysteresis

Makes the reading depend on whether the value is rising or falling

Approach each test point from the same direction

Non-repeatability

Gives slightly different readings for the same value

Average, or accept it

Noise

Adds random variation to each reading

Average, filter, or improve wiring and shielding

Drift

Moves the reading slowly over time or with temperature

Recalibrate on a schedule, and control temperature

Offset, gain and nonlinearity are the same every time, so calibration can remove them. Hysteresis is repeatable too, but it depends on direction. Noise and non-repeatability are random, so averaging reduces them. Drift is slow, so a calibration removes it only until the reading drifts again.

Datasheets state errors in two ways. A percentage of span, the full output range, is a fixed amount across the range. A percentage of reading scales with the value. The NI-9203 current input module, for example, states its gain error as a percentage of reading and its offset error as a percentage of range.[1]

Bits and resolution#

An analog-to-digital converter (ADC) divides its input range into steps. An N-bit ADC has 2N steps, so each step, one least significant bit (LSB), is the range divided by 2N.

The NI-9203 has a 16-bit ADC. In its unipolar mode it measures 0 to 20 mA, with 6.5% of overrange, for a range of 21.5 mA. That gives 330 nA per step.[1]

A 4–20 mA sensor uses only 16 mA of that range, so a 0–100 psig transducer gets about 48,500 steps, or about 0.002 psig per step. The unused steps below 4 mA aren’t wasted: a reading near 0 mA means a broken wire.

Converting to the nearest step adds an error of up to half a step. For an ideal N-bit converter, that sets a best possible signal-to-noise ratio of 6.02 × N + 1.76 dB, or 98 dB for 16 bits.[2] Real converters add their own noise. The NI-9203’s input noise is 1 LSB RMS, and its nonlinearity is up to ±3 LSB.[1]

More bits don’t mean more accuracy. The NI-9203’s calibrated offset error alone is 0.06% of its range, 12.9 µA, which is about 39 steps.[1] Steps smaller than that offset are real changes in the signal, but they don’t say what the true value is.

Sample rate and aliasing#

To capture a signal, sample faster than twice the highest frequency in it. Content above half the sample rate shows up in the data at false, lower frequencies. This is aliasing, and it can’t be undone afterwards, so an anti-aliasing filter goes before the ADC.[3]

The NI-9203 samples at up to 200,000 samples per second,[1] far faster than a pressure transducer can respond: the WIKA A-10’s response time is under 4 ms.[4] On a fluids panel, which runs compressed air through valves and pressure transducers, the pressures change over seconds, so a low sample rate is enough for them.

On an e-axle dyno, where a motor drives the e-axle under test and load motors brake it, the software reads each motor controller about 16 times a second. That’s enough to follow changes slower than 8 Hz, such as speed steps and temperature, but not to measure vibration. Frequency domain works out which of the dyno’s frequencies that rules out, and what an accelerometer would need.

Worked example: an error budget for a pressure reading#

The fluids panel’s supply pressure, p_1, is a 0–100 psig transducer with a 4–20 mA output, read by an NI current input module. This example works out how accurate a reading of 85 psig is, using a real transducer and module:

  • The transducer is a WIKA A-10, 0–100 psig, 4–20 mA, with the 0.25% nonlinearity option. Its accuracy is ±0.5% of span, which includes nonlinearity, hysteresis, offset and gain error. Its long-term drift is ±0.1% of span.[4]

  • The module is an NI-9203 in unipolar mode, calibrated, anywhere from −40 °C to 70 °C. Its maximum gain error is ±0.18% of reading, and its maximum offset error is ±0.06% of its 21.5 mA range.[1]

  • The stand is indoors at 15–25 °C, the transducer’s reference temperature, so the transducer’s temperature error doesn’t apply.[4]

1. Find the signal#

The transducer maps 0–100 psig to 4–20 mA, so 1 mA is 6.25 psig, and 85 psig is:

4 mA + (85 / 100) × 16 mA = 17.6 mA

2. Add up the module’s error#

NI’s formula for a C Series module’s absolute accuracy is the reading times the gain error, plus the range times the offset error, plus noise.[5]

gain:    17.6 mA × 0.18% = 31.7 µA
offset:  21.5 mA × 0.06% = 12.9 µA
noise:   3 × 0.33 µA ÷ √100 ≈ 0.1 µA   (3 sigma, averaging 100 readings)
total:                     ≈ 44.7 µA

At 6.25 psig per mA, 44.7 µA is ±0.28 psig. At the module’s typical accuracy, near 25 °C, the same sum is 11.4 µA, or ±0.07 psig.

3. Add up the transducer’s error#

accuracy:  0.5% × 100 psig = 0.50 psig
drift:     0.1% × 100 psig = 0.10 psig

4. Combine them#

When errors are independent, NI combines them as the square root of the sum of their squares (RSS).[5] The worst case adds them directly:

Source

Error (psig)

Transducer accuracy

±0.50

Transducer long-term drift

±0.10

NI-9203 accuracy, calibrated maximum

±0.28

Combined, RSS

±0.58

Combined, worst case

±0.88

The reading is 85.0 psig, give or take about 0.6 psig.

  • The transducer dominates. A better module wouldn’t help much. A better transducer, or a calibration of the whole chain against a reference gauge, would.

  • The resolution, 0.002 psig, is about 300 times finer than the accuracy. Don’t report more digits than the accuracy supports.

  • Without calibration, the module alone could be off by up to 232 µA, or ±1.45 psig, using the NI-9203’s uncalibrated maximum errors of ±0.66% of reading and ±0.54% of range.[1]

Outside its reference temperature, the transducer’s temperature error can reach 1.0% of span typically and 2.5% at most, ±2.5 psig, over its 0–80 °C operating range.[4] That would dominate every other error in the budget.

The transducer’s signal noise, up to 0.3% of span,[4] is random, so averaging reduces it. Averaging 100 readings divides random noise by √100, a factor of 10.[5]

Uncertainty#

The error budget above gives limits. Metrology states the same result as an uncertainty, using the method in the Guide to the expression of uncertainty in measurement (GUM).[6]

The GUM combines standard uncertainties, which are like standard deviations. To turn a datasheet limit of ±a into a standard uncertainty, treat the true error as equally likely anywhere in that range, which gives a/√3. Combine the standard uncertainties by RSS, then multiply by a coverage factor, k. With k = 2, the result covers about 95% of cases.[7] To estimate a standard uncertainty from repeated readings, see Describing a set of results.

For the example above:

combined standard uncertainty:  0.58 psig ÷ √3 = 0.34 psig
expanded uncertainty (k = 2):   2 × 0.34 psig ≈ 0.7 psig

So the reading is 85.0 psig ± 0.7 psig, at k = 2.

Calibration#

Calibration compares an instrument with a known reference and corrects or records the difference. The NI-9203 ships with a NIST-traceable calibration, and NI recommends recalibrating it every year.[1] Traceable means each reference was itself calibrated, in a chain back to a national standard.

A module calibration covers the module, not the transducer. To calibrate the whole chain, apply known pressures from a reference, record the readings, and fit a correction. That removes the repeatable errors of every part at once.

Checks you can do on a stand#

Each check finds one kind of error from Kinds of error:

  • Zero check. Vent the fluids panel and read p_1. It should read 0 psig. Anything else is offset error.

  • Two-sensor check. Where two sensors measure the same thing, compare them. A difference that grows with the value is gain error.

  • Repeatability check. Run the same step twice and compare the readings.

  • Noise check. Hold a steady condition, and look at the spread of the readings. On the dyno, hold a speed and compare the standard deviation of the speed channel with the speed itself.

  • Hysteresis check. Step a value up, then back down through the same points, and compare.