Fluke8808A#

A driver for InstroDMM (unstable)

Fluke 8808A

The Fluke8808A provides a driver that can be used to instantiate an InstroDMM. It ships in the instro-unstable package (pip install 'instro[unstable]') while its API settles.

The 8808A’s only remote interface is its RS-232 port, which instro reaches as a VISA ASRL resource. The meter speaks a Fluke 45-style command set rather than SCPI. The driver reads the primary display only and supports DC/AC voltage, DC/AC current, 2- and 4-wire resistance, range, and digits.

Behavior to know#

  • Front panel locked while connected. open() resets the meter (*RST) and locks the front panel, so the function the driver tracks can’t change underneath it. close() returns the meter to local control.

  • Ranges apply to the active function. Select the function first (set_measurement_function, or any read_* call), then set its range. Setting a range for another function raises an error. A range value picks the smallest range whose full scale covers it; None selects autorange.

  • Two resolutions. set_digits(5) selects the slow rate (2.5 readings/s); set_digits(4) selects the fast rate (100 readings/s). Other values raise ValueError.

  • Overload. A reading over range (0L on the display) comes back as ±1.0E+9.

Serial settings#

The driver’s serial defaults are 9600 baud, 8 data bits, no parity, 1 stop bit, and no flow control, which match the meter’s factory RS-232 setup. To use other settings, pass a VisaConfig with a matching SerialConfig instead of a resource string.

The driver works with the meter’s Echo setting on or off: on open() it checks whether the meter answers each command with a prompt and adapts to it.

Creating an InstroDMM with Fluke8808A#

from instro.unstable.dmm.drivers import Fluke8808A
from instro import InstroDMM

dmm = InstroDMM(
    name="myDMM",
    driver=Fluke8808A("ASRL/dev/ttyUSB0::INSTR"),
)

Parameters and methods specific to Fluke8808A can be found in the Reference.